Command Set
Commands are non-case sensitive.
Underline means a single space (dc_1→DC 1V).
When the parameter does not match the real value, the closest possible
option is automatically selected (dc_2 [DC 2V range]→DC 10V)
CONFigure command
conf:volt:dc
Sets measurement to DC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:dc_1 (DCV, 1V range)
Example: conf:volt:dc_min (DCV, minimum range)
conf:volt:ac
Sets measurement to AC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:ac_1 (ACV, 1V range)
Example: conf:volt:ac_min (ACV, minimum range)
conf:volt:dcac
Sets measurement to DC+AC Voltage and specifies range.
Parameter: NR2, min, max
Example: conf:volt:dcac_1 (DC+ACV, 1V range)
Example: conf:volt:dcac_min (DC+ACV, minimum range)
conf:curr:dc
Sets measurement to DC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:dc_10e-3 (DCI, 10mA range)
Example: conf:curr:dc_min (DCI, minimum range)
conf:curr:ac
Sets measurement to AC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:ac_10e-2 (ACI, 100mA range)
Example: conf:curr:ac_min (ACI, minimum range)
conf:curr:dcac
Sets measurement to DC+AC Current and specifies range.
Parameter: NR2, min, max
Example: conf:curr:dcac_10 (DC+ACI, 10A range)
Example: conf:curr:dcac_min (DC+ACI, minimum range)
conf:res
Sets measurement to 2W Resistance and specifies range.
Parameter: NR2, min, max
Example: conf:res_10e3 (2W R, 10K range)
Example: conf:res_min (2W R, minimum range)
conf:fres
Sets measurement to 4W Resistance and specifies range.
Parameter: NR2, min, max
Example: conf:fres_10e3 (4W R, 10K range)
Example: conf:fres_min (4W R, minimum range)
conf:freq
Sets measurement to Frequency and specifies range.