Dc current – GW Instek GDM-8200A User Manual User Manual
Page 116
![background image](/manuals/367425/116/background.png)
GDM-8200A Series User Manual
116
F
100.0mV
100µV
120.0mV
199.9mV
1.000V
1mV
1.200V
1.999V
10.00V
10mV
12.00V
19.99V
100.0V
100mV
120.0V
199.9V
750V(*)
1V
750V
750V
Rate
Range
Accuracy (reading%+digits)
20~45Hz
45~10kHz
10k~30kHz 30k~100kHz
S
100.000mV 1% + 100 0.2% + 100 1.5% + 300 5% + 300
1.00000V 1% + 100 0.2% + 100 1% + 100 3% + 200
10.0000V 1% + 100 0.2% + 100 1% + 100 3% + 200
100.000V 1% + 100 0.2% + 100 1% + 100 3% + 200
750.00V(*) 1% + 100 0.2% + 100 1% + 100 3% + 200
M
100.00mV
—
0.2% + 40
1.5% + 80 5% + 120
1.0000V
—
0.2% + 40
1% + 40
3% + 80
10.000V
—
0.2% + 40
1% + 40
3% + 80
100.00V
—
0.2% + 40
1% + 40
3% + 80
750.0V(*)
—
0.2% + 40
1% + 40
3% + 80
F
100.0mV
—
0.2% + 5
1.5% + 10 5% + 15
1.000V
—
0.2% + 5
1% + 5
3% + 10
10.00V
—
0.2% + 5
1% + 5
3% + 10
100.0V
—
0.2% + 5
1% + 5
3% + 10
750V(*)
—
0.2% + 5
1% + 5
3% + 10
DC Current
Note
mA range protected with a 2A fuse
10A range protected with a 12A, 600V fuse
10A only for 30 seconds
Rate
Range
Resolution
Full Scale
GDM-8251A
Full Scale
GDM-8255A
Accuracy
(reading%
+ digits)
S
10.0000mA 0.1µA
12.0000mA 19.9999mA 0.05%+15
100.000mA 1µA
120.000mA 199.999mA 0.05%+5
1.0000A
100µA
1.2000A
1.9999A
0.2%+5
10.0000A 100µA
10.0000A 10.0000A 0.2%+5
M
10.000mA 1µA
12.000mA 19.999mA 0.1%+6
100.00mA 10µA
120.00mA 199.99mA 0.1%+3
1.000A
1mA
1.200A
1.999A
0.2%+3
10.000A
1mA
10.000A
10.000A
0.2%+3
F
10.00mA 10µA
12.00mA 19.99mA 0.1%+2
100.0mA 100µA
120.0mA 199.9mA 0.1%+2
1.00A
10mA
1.20A
1.99A
0.2%+2
10.00A
10mA
10.00A
10.00A
0.2%+2