Application: external trigger – GW Instek GDM-8200A User Manual User Manual

Page 94

background image

GDM-8200A Series User Manual

94

Application: External trigger

Background

The GDM-8200A series uses the internal trigger by
default, for example to count the frequency and the

period. Using an external trigger allows customized
triggering condition.

Signal
connection

Connect the external trigger signal to the Digital I/O

port located on the rear panel.

1 2 3 4 5

6 7 8 9

FAIL Out

HIGH Limit FAIL Out

LOW Limit FAIL Out

EOM Out

VCC Out

NC

Digital (Chassis) Ground

External Trigger In

PASS Out

Pin4

External Trigger Input pin

1. Activate
external trigger

Press the Shift key followed by the
TRIG key. The EXT indicator

appears on the display.

2. Start trigger

Press the TRIG key to start
triggering manually. The

indicator turns On.

Reading indicator

The reading indicator stays On before triggering.
After triggering, the indicator flashes according to the

external signal trigger timing.

Exit external
trigger

Press the Shift key followed by the
TRIG key. The EXT indicator

disappears and the trigger goes back
to internal mode.