:SYSTem:CONFigure:PROTection:RECovery

Set

Query

Description

Sets or queries how the OHP, FAN, AC-FAIL, and
SD alarms are cleared.

Syntax
Return Syntax

:SYSTem:CONFigure:PROTection:RECovery
{SAFE|AUTO}
:SYSTem:CONFigure:PROTection:RECovery?

Parameter

SAFE

The output is not turned on automatically
when the cause of the alarm is fixed.

AUTO

The output is turned on automatically
when the cause of the alarm is fixed.

:SYSTem:CONFigure:MSLave

Set

Query

Description

Sets or queries the unit operation mode. This
setting is only applied after the unit has been reset.

Syntax
Query Syntax

:SYSTem:CONFigure:MSLave { }
:SYSTem:CONFigure:MSLave?

Parameter/Return Description

0

Master/Local

1

Master/with 1 unit in Parallel (total:2 units)

2

Master/with 2 units in Parallel (total: 3 units)

3

Master/with 3 units in Parallel (total: 4 units)

4

Slave

:SYSTem:CONFigure:OUTPut:EXTernal
[:MODE]

Set

Query

Description

Sets the external logic as active high or active low.
This setting is only applied after the unit has been
reset.

Syntax
Query Syntax

:SYSTem:CONFigure:OUTPut:EXTernal[:MODE]
{|LOW|HIGH}
:SYSTem:CONFigure:OUTPut:EXTernal[:MODE]?