Initiate Commands

:INITiate:CONTinuous[:TRANsient] ................................ 64

:INITiate[:IMMediate]:NAME ............................................ 64

:INITiate[:IMMediate][:TRANsient] ................................... 65

:INITiate:CONTinuous[:TRANsient]

Set

Query

Description

This command continuously initiates software
triggers for the transient or output triggers.

Syntax
Query Syntax

:INITiate:CONTinuous[:TRANsient] {|OFF|ON}
:INITiate:CONTinuous[:TRANsient]?

Parameter

OFF | 0

OFF

ON | 1

ON

Return parameter 0

OFF

1

ON

Example

INIT:TRAN 1
Turns on the continuous trigger.

:INITiate[:IMMediate]:NAME

Set

Description

The INITiate command starts the TRANsient or
OUTPut trigger.

Syntax

:INITiate[:IMMediate]:NAME {TRANsient|OUTPut}

Parameter

TRANSient

Starts the TRANsient trigger.

OUTPut

Starts the OUTPut trigger.

Example

INITiate:NAME TRANient
Starts the TRANSient trigger.