beautypg.com

Source]:voltage[0]:protection[:level, Source]:voltage[0]:protection:tripped – GW Instek PSU-Series Programming User Manual User Manual

Page 85

background image

REMOTE CONTROL

85

[:SOURce]:VOLTage[0]:PROTection[:LEVel]

Set

Query

Description

Sets or queries the overvoltage protection level.

Syntax

[:SOURce]:VOLTage[0]:PROTection[:LEVel]
{(V)|MINimum|MAXimum}

Query Syntax

[:SOURce]:VOLTage[0]:PROTection[:LEVel]?

Parameter/Return

Minimum: Depends on the unit type:
if Vrated * 0.1 > 5V, then Minimum = 5V,
else Minimum = Vrated * 0.1
Maximum: Vrated * 1.1

MIN

Minimum OVP level

MAX

Maximum OVP level

Example

SOUR:VOLT:PROT:LEV MAX
Sets the OVP level to its maximum.

[:SOURce]:VOLTage[0]:PROTection:TRIPped

Set

Query

Description

Sets or queries the overvoltage protection level.

Query Syntax

[:SOURce]:VOLTage[0]:PROTection:TRIPped?

Return parameter

0

Protection not tripped

1

Protection tripped

Example

SOUR:VOLT:PROT:TRIP?
>0
Indicates that the OVP protection has not been
tripped.