[:SOURce]:VOLTage[0]:LIMit:AUTO

Set

Query

Description

Sets whether to limit the voltage setting so that it
does not exceed the OVP setting or become lower
than the UVP setting.
If you enable the limit when the OVP setting is
lower than the voltage setting, the OVP setting will
be set to 105 % of the voltage setting.
If you enable the limit when the UVP setting is
higher than the voltage setting, the UVP setting
will be set equal to the voltage setting.

Syntax

[:SOURce]:VOLTage[0]:LIMit:AUTO {|OFF|ON}

Query Syntax

[:SOURce]:VOLTage[0]:LIMit:AUTO?

Parameter

OFF | 0

Disable the limit setting

ON | 1

Enable the limit setting

Return parameter

Returns the setting in format.

Example

SOUR:VOLT:LIM:AUTO 0
Disables the limit setting.

[:SOURce]:VOLTage[0]:LIMit:LOW

Set

Query

Description

Sets or queries the under voltage (UVL) trip point.

Syntax

[:SOURce]:VOLTage[0]:LIMit:LOW
(V)|MINimum|MAXimum

Query Syntax

[:SOURce]:VOLTage[0]:LIMit:LOW?

Parameter/Return

0 ~ the present setting voltage

MIN

Minimum allowed voltage level

MAX

Maximum allowed voltage level

Example

SOUR:VOLT:LIM:LOW MAX
Sets the UV> level to its maximum.