PSU Programming Manual

44

F-76 = 00~31

Set the address of each slave to
a unique address identifier

7. Multiple units can now be operated at the same

time, see the function check below for usage
details.

Serial link cable
with RJ-45
shielded
connectors from
PSU-232 or PSU-
485 connection
kit

8 Pin Connector (IN)

8 Pin Connector (OUT)

Pin No.

Name

Pin No.

Name

Housing

Shield

Housing

Shield

1

SG

1

SG

6

TXD -

6

TXD -

3

TXD +

3

TXD +

5

RXD -

5

RXD -

4

RXD +

4

RXD +

Multi-Unit Function Check

Functionality
check

Invoke a terminal application such as Realterm.

To check the COM port No, see the Device
Manager in the PC. For WinXP; Control panel
→ System → Hardware tab.

Below shows an example using the TDK
GENESYS legacy commands.

TDK Query

When using the TDK GENESYS legacy
commands, each unit can be individually
controlled using the unique address identifiers.
For this function check, we will assume that the
master unit is assigned to address 8, while a
slave is assigned address 11.