Source]:current[:level]:triggered [:amplitude, Source]:current:limit:auto, Source]:current:protection:delay – GW Instek PSU-Series Programming User Manual User Manual
Page 78

PSU Programming Manual
78
[:SOURce]:CURRent[:LEVel]:TRIGgered
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps when a
software trigger has been generated.
Syntax
[:SOURce]:CURRent[:LEVel]:TRIGgered[:AMPLitude]
{
Query Syntax
[:SOURce]:CURRent[:LEVel]:TRIGgered[:AMPLitude]?
Parameter
0%~105% of the rated current output in amps.
MIN
Minimum current level.
MAX
Maximum current level.
Return Parameter
Returns the current level.
Example
SOUR:CURR:LEV:TRIG:AMPL?
38.000
Returns the maximum possible current level in amps.
[:SOURce]:CURRent:LIMit:AUTO
Set
Query
Description
Enables or disables the limit on the current setting.
Syntax
[:SOURce]:CURRent:LIMit:AUTO {
Query Syntax
[:SOURce]:CURRent:LIMit:AUTO?
Parameter
OFF | 0
Disable the setting current limit
ON | 1
Enable the setting current limit
Return parameter
Returns the setting in
Example
SOUR:CURR:LIM:AUTO 0
Disables the current limit.
[:SOURce]:CURRent:PROTection:DELay
Set
Query
Description
Sets the Delay Time for OCP in seconds for turning
the output off. The delay is set to 0.1 by default.
Syntax
[:SOURce]:CURRent:PROTection:DELay
{