GW Instek PSU-Series Programming User Manual User Manual

Page 39

background image

REMOTE CONTROL

39

PSU-485 RS485
cable with DB9
connector

DB-9 Connector

Remote IN Port

Remarks

Pin No.

Name

Pin No.

Name

Housing Shield

Housing Shield

9

TXD -

6

RXD -

Twisted
pair

8

TXD +

3

RXD +

1

SG

1

SG

5

RXD -

5

TXD -

Twisted
pair

4

RXD +

4

TXD +

Steps

1. Connect the RS232 serial cable

(included in the PSU-232 kit) or
RS485 serial cable (included in the
PSU-485 kit) to the (Remote IN
port) on the rear panel.

Connect the other end of the cable
to the PC.

RS 485 / 232

2. Press the Function key to enter the

Normal configuration settings.

Page 21

Set the following UART settings:

F-70 = 1 or 2

Interface: 0= Disable UART,
1= RS232 or 2 = RS485

F-71 = 0 ~ 7

Set the baud rate:
0=1200, 1=2400, 2=4800,
3=9600, 4=19200, 5=38400,
6=57600, 7=115200

F-72 = 0 or 1

Data bits: 0=7 or 1=8

F-73 = 0 ~3

Parity: 0 = none, 1 = odd, 2 =
even

F-74 = 0 or 1

Stop bits: 0 = 1, 1 = 2

F-75 = 0 or 1

TCP: 0 = SCPI, 1 = TDK
(emulation mode)