Output:pon[:state, Output[:state][:immediate – GW Instek PSU-Series Programming User Manual User Manual
Page 70
![background image](/manuals/367357/70/background.png)
PSU Programming Manual
70
:OUTPut:PON[:STATe]
Set
Query
Description
Sets the output state at power-on. This is the
equivalent to the F-92 (Output Status when Power
ON) power on configuration settings. These
settings only apply after the unit has been reset.
Syntax
Return Syntax
:OUTPut:PON[:STATe] {
:OUTPut:PON[:STATe]?
Parameter
SAFE | 0
The PSU turns on in the same state the unit
was in prior to the previous shut down. The
output is set to off (default).
FORCe | 1
The PSU turns on in the same state the unit
was in prior to the previous shut down. The
output is set to on.
AUTO | 2
The PSU turns on in the same state the unit
was in prior to the previous shut down, but
with the same output on/off setting.
Return parameter 0
The power on output setting is “SAFE”.
1
The power on output setting is “FORCe”.
2
The power on output setting is “AUTO”.
:OUTPut[:STATe][:IMMediate]
Set
Query
Description
Turns the output on or off.
Syntax
Query Syntax
:OUTPut[:STATe][:IMMediate] {
:OUTPut[:STATe][:IMMediate]?
Parameter
OFF | 0
Turns the output off.
ON | 1
Turns the output on.
Return parameter
Returns output status of the instrument.