beautypg.com

Source]:voltage[0][:level][:immediate] [:amplitude, Source]:voltage[0][:level]:triggered [:amplitude – GW Instek PSU-Series Programming User Manual User Manual

Page 83

background image

REMOTE CONTROL

83

Example

VOLT:EXT:SOUR?
NONE
External control is not used.

[:SOURce]:VOLTage[0][:LEVel][:IMMediate]
[:AMPLitude]

Set

Query

Description

Sets or queries the voltage level in volts.

Syntax

[:SOURce]:VOLTage[0][:LEVel][:IMMediate][:AMPLitude
] {(V)|MINimum|MAXimum}

Query Syntax

[:SOURce]:VOLTage[0][:LEVel][:IMMediate][:AMPLitude
]?

Parameter

0~105% of the rated output voltage in volts.

MIN

Minimum voltage level

MAX

Maximum voltage level

Return parameter

Returns the voltage level in volts

Example

SOUR:VOLT:LEV:IMM:AMPL 10
Sets the voltage level to 10 volts.

[:SOURce]:VOLTage[0][:LEVel]:TRIGgered
[:AMPLitude]

Set

Query

Description

Sets or queries the voltage level in volts when a
software trigger has been generated.

Syntax

[:SOURce]:VOLTage[0][:LEVel]:TRIGgered[:AMPLitude]
{(V)|MINimum|MAXimum}

Query Syntax

[:SOURce]:VOLTage[0][:LEVel]:TRIGgered[:AMPLitude]?

Parameter

0%~105% of the rated voltage output in volts.

MIN

Minimum current level.

MAX

Maximum current level.

Return parameter

Returns the voltage level.

Example

SOUR:VOLT:LEV:TRIG:AMPL 10
Sets the voltage level to 10 volts when a software