PSU Programming Manual

80

Parameter

OFF | 0

Turns the OCP off.

ON | 1

Turns the OCP on.

Return parameter

Returns the over current protection state in

format.

Example

SOUR:CURR:PROT:STAT OFF
Turns OCP off.

[:SOURce]:CURRent:PROTection:TRIPped

Set

Query

Description

Returns the state of the current protection circuits.

Query Syntax

[:SOURce]:CURRent:PROTection:TRIPped?

Return parameter

Returns protection status.

Example

SOUR:CURR:PROT:TRIP?
>0
The protection circuit has not been tripped.

[:SOURce]:CURRent:SLEW:RISing

Set

Query

Description

Sets or queries the rising current slew rate. This is
only applicable for CC slew rate priority mode.

Syntax

Query Syntax

[:SOURce]:CURRent:SLEW:RISing
{(A)|MINimum|MAXimum}
[:SOURce]:CURRent:SLEW:RISing?

Parameter

Per step is Rated Current divided by 100 ms.

MIN

Minimum rising current slew rate.

MAX

Maximum rising current slew rate.

Return parameter

Returns the step current in amps.

Example

SOUR:CURR:SLEW:RIS?
0.950
Sets the rising current slew rate to 0.950 A/ms.