M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 82

M Series Service Manual
72
SYSTEM FAULT 5
ECU RAM test failure or ROM checksum test
failure.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
SYSTEM FAULT 6
No communications received from ECU for 4
seconds.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
SYSTEM FAULT 7
The A/D converter is not performing conversions in
a timely manner.
Replace system board.
SYSTEM FAULT 36
P1MON is less than 412 counts or greater than 612
A/D counts. Pace/defib is disabled as long as
condition exits.
Replace system board.
SYSTEM FAULT 37
Disable pace/defib and MFE monitoring.
Replace system board.
SYSTEM FAULT 38
Failure to shutdown after “shutdown order” is
written to the RTC.
Replace system board.
TEST FAILED
MCU performed ipeak test (defib peak current) and
unit failed during 30J self test.
Replace universal
cable, paddles or high
voltage module,
capacitor, or system
board.
TEST OK
MCU performed ipeak test (defib peak current) and
unit passed 30J self test.
USE PADDLE DISCHG
Front Panel discharge button is pressed when either
external paddles or internal spoons with discharge
buttons are connected.
Error Message
Explanation
User
Advisory
Technical Action