M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 32

M Series Service Manual
22
10.9
Remove an EEG lead wire from
the QED 6 analyzer.
The alarm symbol flashes and the heart symbol stops
flashing. The ECG LEAD OFF alarm tone sounds.
Recorder prints a stripchart showing a low heart rate, if
enabled.
o
o
o
10.10
Reattach ECG Lead wire to the
QED 6 Analyzer and hold the
ALARM SUSPEND button on
unit for 4 seconds.
The unit displays the alarm symbol with an “X” through
it.
The heart symbol flashes with each QRS wave.
o
o
o
10.11
Press the ALARM SUSPEND
button to enable the alarms.
The alarm symbol does not have an “X” through it.
o
o
o
10.12
Set QED 6 Analyzer to 160 BPM
or higher.
The Heart Rate Value is highlighted, the alarm tone
sounds, and the alarm and the heart symbol both flash.
o
o
o
10.13
Press the ALARM SUSPEND
button on the M Series unit.
The alarm is suspended for 90 seconds. The unit
displays the alarm symbol with an “X” through it, and
the heart symbol flashes with each QRS wave.
Note: Software version 35.00-36.XX, the alarm is
suspended until the alarm button is press again, or the
fault is cleared.
o
o
o
10.14
Press and hold ALARM
SUSPEND for 4 seconds to
disable alarms.
o
o
o
Record your results on the Maintenance Tests Checklist.
Do this...
Verify that...
Pass Fail/
NA