ZOLL M Series Defibrillator Rev R User Manual
Page 13

M Series Service Manual
3
Equipment that You Need to Perform the M Series Maintenance
Tests
This section lists equipment that we use to perform the maintenance tests that we describe in this chapter. You can
substitute an equivalent device for a listed device; however, not all simulators and analyzers will produce the same
results. Be sure to follow the manufacturer’s recommendations for conducting the maintenance tests.
We recommend the use of the following equipment when performing M Series Maintenance Tests:
• ZOLL Medical Electrode Adapter from Fluke Biomedical (ZOLL Part Number 3010-0378).
• QED 6 Defibrillator Analyzer
• MFC Test Port Connector, 1004-0053-99
• Fluke Biomedical
601 Pro Series International Safety Analyzer.
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).
• Stop watch.
• Standard series II PC flash memory cards.
• 1 red miniature alligator to miniature alligator test lead.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supply (15 amp minimum).
• 0.1 ohm 1% resistor (1/4 watt or greater).
• 1000 ohm 1% resistor (1/4 watt or greater).
• Fluke 75 Multimeter or equivalent.
• Printer Paper.
• Battery.
• AC line cord.
• 3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.