M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 80

M Series Service Manual
70
REPLACE EtCO
2
SENSOR
EtC0
2
SENSOR WARM UP message displays for
more than five minutes. Sensor defective.
Replace sensor cable.
REPORT FULL
Summary report memory full.
Erase summary report.
REPORT HALTED
Summary report stops printing unexpectedly.
Turn unit off and then
back on again. Print
Summary again. If fault
persists, replace system
board.
RESERVED 1
The watchdog timer is not functional in the unit.
Turn off unit and then
turn on to reset. If fault
persists, replace system
board.
RETRY ANALYSIS
Advisory message in conjunction with noisy ECG.
Analysis halted.
SELECT 30J FOR TEST
Attempt to run a self test at an energy other than 30J.
SELECT DEFIB MODE
Analyze button pressed in pace or monitor mode.
SELECT LIMB LEADS
Paddles or augmented ECG leads selected when
continuous analysis active or started.
Select limb leads I, II,
III or MFE
SELECT PADS
Lead I, II, or III selected when analyze pressed.
SET CLOCK
Real time clock failure: invalid date or time.
Set date and time and/
or replace system
board.
SET PACE mA
Multiple copy errors are the product of intended
software or memory errors. If error reoccurs other
than on entering pace the first time or after more than
10 minutes in other mode, the unit could be broken.
Set pace current. If
broken, replace system
board.
Error Message
Explanation
User
Advisory
Technical Action