M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 71

M Series Service Manual
61
CARD FULL
Memory Card Full.
CHECK CO
2
SENSOR
EtCO
2
Sensor is unplugged or defective.
Check that sensor cable
is plugged in and seated
properly. Check that
sensor is not exposed to
excessive heat. If
problem persists,
replace the sensor.
CHECK CO
2
ADAPTER
Airway adapter is removed, occluded or adapter
zeroing needs to be performed or was performed
incorrectly.
Replace/Clean airway
adapter. Zeroing
performed
automatically.
CHECK MEMORY CARD
No card detected during manual or semi-automated
modes.
CHECK PADS
Message displayed in conjunction with either POOR
PAD CONTACT or DEFIB PAD SHORT.
Ensure pads are
coupled to patient.
Check /replace pads
and universal cable.
Replace system board.
CHECK PATIENT
Background ECG analysis detects shockable rhythm.
CHECK PULSE
Alternate message for NO SHOCK ADVISED
message.Message also shown after delivering third
shock when auto analyze 3 times option is enabled.
CHECK RECORDER
Produced when paper tray is empty, paper jams or
recorder door is opened.
Replace paper sensor
board, system
interconnect board,
and/or system board.
Error Message
Explanation
User
Advisory
Technical Action