PRESS CHARGE
Advisory message in conjunction with shock
advised.
PRESS SHOCK
Prompt issued in AED auto defib mode when defib is
charged (ready).
RECORDER FAULT 142
Strip chart system error.
Check paper tray and
paper path. Replace the
print head, system
interconnect board and
or the system board.
RECORDER FAULT 143
Strip chart failed power-up echo test.
Communications error.
Check paper tray and
paper path. Replace the
system interconnect
board and/or the system
board. Turn unit off and
back on again.
RECORDER FAULT 147
Strip chart printhead over temperature.
Check paper tray and
paper path. Replace the
print head, system
interconnect board and/
or the system board.
RELEASE BUTTONS
Simultaneous external paddle button presses
detected before unit reached full defib charge (ready
state).
Release buttons.
RELEASE SHOCK
Discharge switch(es) closed when pressing charge
button. Discharge button pressed before defib
reached ready state.
Release shock button.
Check paddles.
Replace controls board.
REPLACE BATTERY
Battery voltage is less than absolute minimum.
Shutdown imminent.
Replace with charged
battery.