M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 78

M Series Service Manual
68
PACER FAULT 123
Measured pace rate is too fast compared to selected
rate.
Replace high voltage
module or system
board.
PACER FAULT 126
Issued in conjunction with message 122. Pace
current is more than 15mA and below selected value.
Replace high voltage
module.
PADDLE FAULT
Cannot detect type of accessory attached to the
universal cable.
Replace paddles,
internal paddles, system
board, high voltage
module and/or
universal cable.
PERFORM CPR
Advisory message in AED auto defib mode.
PLACE ON ZERO CELL
EtCO
2
sensor cable plugged into unit for the first
time. Zeroing error or probe drift error detected.
Zero sensor. Replace
sensor. Return to ZOLL
for service.
POOR LEAD CONTACT
One or more ECG leads are poorly connected or not
connected to patient. (User configurable.)
Check electrode
attachment to patient,
cable connector to
electrode, cable to unit
connector. Broken unit.
POOR PAD CONTACT
Electrode impedance exceeds threshold.
Ensure pads are
coupled to patient.
Check /replace pads or
universal cable.
Check impedance
circuit calibration.
Replace system board.
PRESS ANALYZE
Alternate message for check patient prompt.
Error Message
Explanation
User
Advisory
Technical Action