M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 67

M Series Service Manual
57
Displayed HR not accurate;
artifact present.
•
Reduce or eliminate ECG artifact due to
electrode or patient cable movement. Route
cables so that they don’t pull on electrodes or
swing excessively.
•
Ensure patient is motionless.
•
Check for possible excessive radio frequency
interference.
•
Verify a good connection of electrodes to the
patient.
•
Prepare the patient’s skin prior to the electrode
attachment.
•
Move patient cables away from other electrical
equipment, especially any RFI source.
•
Ensure ECG cable fits snugly in unit.
•
Change ECG cable.
•
Replace/reposition ECG electrodes.
•
Check for contamination on snaps.
Ensure springs are intact.
•
Check for intermittent ECG patient
cable or connector wiring.
•
Replace ECG input connector.
•
Replace ECG connector to the
system board cable.
•
Replace system board.
Wandering baseline.
See “Displayed HR not accurate.” above. Note that
in 90% of electrode issues, size and lead changes
don’t help.
Same as above example.
Electronic interference.
Check for possible excessive radio frequency
interference.
Move patient cables away from other electrical
equipment.
•
Turn off sources of excessive RFI.
•
Move M Series unit away from
RFI source.
Reported Problem
Recommended User Action
Recommended Technical Action