M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 50

M Series Service Manual
40
20.9
On the Patient Simulator, set BP 
channel 1 to an atmospheric 
pressure of 0 mmHg (waveform 
10) and press EXECUTE. Set BP 
channel 2 to duplicate BP channel 
1 (waveform 22) and press 
EXECUTE.
20.10
On the unit go to the IBP menu 
(press the Param softkey and 
select the IBP option), then press 
the Zero softkey to display the Zero 
menu, select the Zero Both option 
and press the Enter softkey.
Verify the unit zeroes both pressure channels and 
displays the P1 ZEROED and P2 ZEROED messages 
within 15 seconds.
Verify that the values displayed for the P1 and ART 
pressure channels are accurate within ±3 mmHg of 
atmospheric pressure (that is, 0 mmHg):
systolic pressure: -3 to 3 mmHg
diastolic pressure: -3 to 3 mmHg
mean pressure: -3 to 3 mmHg
Note: The unit displays pressure values for each pressure 
channel within 15 seconds.
o
o
o
o
o
o
20.11
On the Patient Simulator, set to an 
arterial pressure of 120/80 mmHg 
(waveform 00) and press 
EXECUTE.
Verify the values displayed for the P1 and ART pressure 
channels:
systolic pressure: 116 - 124 mmHg
diastolic pressure: 77 - 83 mmHg
mean pressure: 90 - 98 mmHg
Note: The unit displays pressure values for each pressure 
channel within 15 seconds.
o
o
o
Do this...
Verify that...
Pass/Fail/
NA
