On the Pressure Transducer
Simulator, set the pressure polarity
to positive, and select a pressure
of 500 mmHg.
On the unit, verify that the value displayed for the ART P2
pressure channel is in the range of 291 - 300 mmHg.
(Note: value will be displayed within 15 seconds.)
o
o
o
20.34
Connect the Pressure Transducer
Simulator to the first IBP channel
connector (P1). On the Pressure
Transducer Simulator, set the
pressure polarity to negative, and
select a pressure of -200 mmHg.
20.35
Zero the pressure channel.
The unit calibrates the pressure channel against a
negative pressure of -200 mmHg and displays the P1
ZEROED message.
o
o
o
20.36
On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.
On the unit, verify that the value displayed for the P1
pressure channel is in the range of -47 to -50 mmHg.
(Note: value will be displayed within 15 seconds.)
o
o
o
20.37
Connect the Pressure Transducer
Simulator to the second IBP
channel connector (P2). On the
Pressure Transducer Simulator, set
the pressure polarity to negative,
and select a pressure of -200
mmHg.
20.38
Zero the pressure channel.
The M Series CCT unit calibrates the pressure channel
against a negative pressure of -200 mmHg and displays
the P2 ZEROED message.
o
o
o
20.39
On the Pressure Transducer
Simulator, select a negative
pressure of -250 mmHg.
Verify that the value displayed for the P2 pressure
channel is in the range of -47 to -50 mmHg. Note: Value
will be displayed within 15 seconds.
o
o
o
Do this...
Verify that...
Pass/Fail/
NA