M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 77

M Series Service Manual
67
NO SHOCK ADV
No shock advised. Advisory message when analysis
finds non-shockable rhythm.
NOISY ECG
Number of noisy analysis intervals exceeds
threshold.
Stop all patient
movement. Check
connections. Press
Analyze button again.
OPEN AIR DISCHARGE
Cap voltage too high after discharge attempt, e.g.,
full energy discharge did not occur.
Replace paddles, and,
or high voltage module
and system board.
PACER DISABLED
User prompt issued simultaneously with other pace
faults if pacing is disabled.
Replace high voltage
module or system
board.
PACER FAULT 115
Flyback pulse width control circuit is not under
proper control of the processor and gate array.
Replace high voltage
module, capacitor, or
system board.
PACER FAULT 116
Failure to detect XPACE_ON.
Replace high voltage
module, capacitor, and/
or system board.
PACER FAULT 117
Pace relay is stuck closed.
Replace high voltage
module, capacitor, and/
or system board.
PACER FAULT 121
During pace, the pace pulse width <30ms or >50ms.
Replace high voltage
module, or system
board.
PACER FAULT 122
Pace current is more than 15mA above and below
selected value.
Replace high voltage
module, or system
board.
Error Message
Explanation
User
Advisory
Technical Action