Page 44

M Series Service Manual
34
17.4
Using the Index SpO
2
Simulator,
press the BPM+ or BPM- softkey
until the heart rate is 230 BPM.
The SpO
2
simulator screen displays an SpO
2
rate of 230
BPM.
Note that you may need to wait up to 2 minutes for the
information to appear on the ZOLL display.
The M Series monitor displays an SpO
2
saturation of 96-
100%.
The M Series monitor displays a heart
rate of 226-234
BPM.
o
o
17.5
Using the Index SpO
2
Simulator,
press the BPM- softkey until the
heart rate is 50 BPM.
The SpO
2
simulator screen displays an SpO
2
saturation
of 96-100%.
The M Series monitor displays a heart rate of 46-54
BPM.
o
o
o
o
17.6
Using the Index SpO
2
Simulator,
press the 02+ softkey until the
SpO
2
output is at 72%.
The SpO
2
simulator screen displays an SpO
2
saturation
of 70-74%.
The M Series monitor displays a heart rate of 46-54
BPM.
o
o
o
o
17.7
Press Wave 2 softkey.
Select the SpO2 waveform.
For CCT units: Press the Traces
softkey, and then select SET
TRACE 2, and press Enter.
Select SPO2, press Enter.
Plethysmographic waveform appears on the ZOLL
display.
o
o
17.8
Press RECORDER.
The plethysmographic waveform prints on the strip chart
paper.
o
o
17.9
Using the Index SpO
2
Simulator,
press the BPM- softkey until the
heart rate is at 230 BPM.
The SpO
2
simulator screen displays an SpO
2
saturation
rate of 70-74%.
The M Series monitor displays a heart rate in the heart
position of
226-234 BPM.
o
o
17.10
Remove the Masimo
patient
cable.
Do this...
Verify that...
Pass/