M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 73

M Series Service Manual
63
DEFIB FAULT 72
General defib error.
Turn the unit to OFF
and back on. If fault
persists, replace high
voltage module.
DEFIB FAULT 76
Capacitor voltage too high for selected energy.
Replace high voltage
module or capacitor.
DEFIB FAULT 77
Capacitor voltage > than absolute rated max.
Replace high voltage
module or capacitor.
DEFIB FAULT 78
Unable to charge defib cap.
Replace high voltage
module or capacitor.
DEFIB FAULT 79
Defibrillator charging too slowly.
Replace high voltage
module or capacitor.
DEFIB FAULT 80
4 defibrillator faults detected within 20 second
period.
Replace high voltage
module or capacitor.
DEFIB FAULT 81
Discharge switch in undefined state.
Replace high voltage
module or capacitor.
DEFIB FAULT 84
“Upper” discharge transistor shorted (measured via
applicable A/D channel).
Replace high voltage
module.
DEFIB FAULT 85
“Lower” discharge transistor shorted (measured via
applicable A/D channel).
Replace high voltage
module.
DEFIB FAULT 86
One discharge switch closed during power up test.
Replace paddles,
control board or system
board.
DEFIB FAULT 87
Both discharge switches closed during power up test.
Replace paddles,
control board or system
board.
Error Message
Explanation
User
Advisory
Technical Action