M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 74

M Series Service Manual
64
DEFIB FAULT 94
Processor fault causing safety monitor port to be
non-functional.
Replace system board,
high voltage module or
capacitor.
DEFIB FAULT 95
Safe or shutdown line is not functional.
Replace high voltage
module.
DEFIB FAULT 96
XPATREL or XPAT_ENABLE is faulted or one of
the discharge transistors has shorted.
Replace high voltage
module.
DEFIB FAULT 108
VMON voltage is less than the target energy during
charging.
Replace high voltage
module or capacitor.
DEFIB FAULT 109
Defib capacitor voltage is greater than selected
energy when defibrillator is charging or ready.
Replace high voltage
module or capacitor.
DEFIB FAULT 111
Defib capacitor voltage has exceeded the absolute
maximum acceptable voltage.
Replace high voltage
module, capacitor, and
or system board.
DEFIB NOT CHARGED
Discharge button is pressed but the unit is not
charged.
DEFIB PAD SHORT
Measured impedance between high voltage leads of
MFC.
Ensure pads are
coupled to patient.
Check /replace pads or
universal cable.
Replace system board.
DISABLE SYNC
Sync mode active when analyze pressed in defib.
DISCHARGE FAULT
Defib capacitor voltage is not decreasing.
Replace high voltage
module, capacitor, and/
or system board.
Error Message
Explanation
User
Advisory
Technical Action