Before you begin the maintenance tests, M series service manual – ZOLL M Series Defibrillator Rev R User Manual
Page 12

M Series Service Manual
2
11.
Defibrillator Self Test
12.
Synchronized Cardioversion
13.
Shock
14.
Summary Report
15.
Advisory Message
16.
Pacer
17.
SpO
2
Monitor
18.
EtCO
2
Monitor
19.
Temperature Test
20.
IBP Monitoring Test
21.
NIBP Transducer Calibration Test
22.
23.
NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer
24.
Before You Begin the Maintenance Tests
• Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance
Tests” section shown below.
• Keep an extra fully charged ZOLL M Series compatible battery available.
• Schedule an hour to conduct the entire maintenance test.
• Photocopy the checklist at the back of this document and use the copy to record your results. As you conduct each
step of a procedure, mark the
Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.