Aux Output
In addition to a calibration signal, the following signals can be output through
the AUX OUTPUT connector:
Square Wave
Trigger Out - can be used to trigger another
oscilloscope
DC level - a reference level
Trigger Enabled - can be used as a gating
function to trigger another instrument when the
oscilloscope is ready
Pass/Fail - allows you to set a pulse duration
from 1 ms to 500 ms. This pulse occurs when
pass/fail testing is active and conditions are met.
Aux Output Off - turns off the auxiliary output
signal
Auxiliary Output Setup
Touch Utilities Utilities Setup... from the menu bar.
Touch the Aux Output tab.
If you simply want a 1 kHz, 1 V square wave, touch the button so labeled.
If you want a specialized output, touch one of the buttons under Use Auxiliary
Output For.
Touch inside the Amplitude field and enter a value, using the pop-up numeric
keypad. If you want a TTL level signal, touch the TTL Level checkbox. The
Amplitude field becomes unavailable.
If you selected Square Wave, touch inside the Frequency field and enter a
value, using the pop-up keypad. You can set a value from 5.0 Hz to 5 MHz.
If you selected Pass/Fail, touch inside the Pulse Duration field and enter a
value from 1 ms to 500 ms, using the pop-up numeric keypad.