beautypg.com

Coupling, Deskew, Probe attenuation – Teledyne LeCroy WaveRunner МXi-A Getting Started Manual User Manual

Page 33

background image

Getting Started Manual

922138-00 Rev A

25

B

At this point, the following dialog is shown at the bottom of the display, and
the grid area shortens until this dialog is closed (by touching the Close button
on the upper right hand corner of this dialog).

Coupling

Coupling choices are as follows:

DC 50 Ω

GROUND

DC 1 MΩ

AC 1 MΩ

Select coupling by touching inside the Coupling field and choosing a coupling
mode from the pop-up menu.

NOTE: The coupling choices for a channel change if a ProBus probe is
connected to a channel.

Deskew

Deskew allows you to compensate for different lengths of cables, probes, or
anything else that might cause timing mismatches between signals. Connect
all probes to the desired channels, then probe a common signal with each
probe and adjust for timing differences using Deskew.

Probe Attenuation

If you use a Teledyne LeCroy ProBus compatible active probe, or a probe
compatible with Probe Ring, the attenuation is automatically set by the
oscilloscope. If it is not automatically set, select a value here by touching
inside the Probe Atten field selecting a value from the pop-up menu.

This manual is related to the following products: