Getting started manual – Teledyne LeCroy WaveRunner МXi-A Getting Started Manual User Manual
Page 89

Getting Started Manual
922138-00 Rev A
81
B
Touch inside the Compare Values field and select All or Any from
the pop-up menu.
By selecting All, the test is true only if every waveform falls within
the set limit. By selecting Any, the test is true if just one waveform
falls within the limit.
Touch inside the Condition field in the ParamCompare mini-dialog and select
a math operator from the pop-up menu.
Touch inside the Limit field and enter a value, using the pop-up numeric
keypad. This value takes the dimensions of the parameter that you are testing.
For example, if you are testing a time parameter, the unit is seconds. If you
chose either WithinDeltaPct or WithinDeltaAbs from the Condition menu,
you also have the choice of setting the limit by means of the statistical buttons
at the bottom of the ParamCompare dialog: