PLEASE NOTE THE FOLLOWING:

You can tune the dwell time that the scope will wait before loading
the next trigger setup using the Dwell Time data entry field.

If you have Persistence display mode enabled, all trigger events are
recorded on the display. Refer to the Persistence Setup topic for
instructions on enabling Persistence display mode.

If you want TriggerScan to stop when the scope triggers next, check the Stop
On Trigger checkbox. You can use this to isolate specific trigger setups.

Saving TriggerScan Setups

You should save TriggerScan setups once you have made any modifications to
the Trigger List. The current Trigger List will not be preserved after exiting the
application unless you save it.

On the TriggerScan dialog, touch inside the Setup File Name data entry field
and enter a file name using the pop-up keypad.

OR

Touch the Browse button and select a location and file name.

Finally, make sure you touch the Save Setup button.

NOTE: You can load previously saved TriggerScan setups by touching the
Browse button, locating the file, and then clicking Load Setup.