Section 5: Input/ Output

07/2012

87

51254 Rev. 3

5.7.3. SERIAL 20 MA CONFIGURATION (SERIAL

EXPANSION BOARD)

1. Select

LOGIN

.

2. Enter the Supervisor Password or

Service Password.

3. Select the

CONFIGURATION

MENU

.

4. Press PAGE FORWARD once.

5. Select

REMOTE DISPLAY

.

6. Touch the data box on the screen to

the right of

COM PORT:

, selecting

the desired COM port.

7. Select the

DISPLAY MODE

button,

selecting the desired mode required
from the menu list.

– The menu choices are Continuous

Display or Display on Print.

8. Touch the data box on the screen to the right of

OUTPUT TYPE

:, then

select the desired type required from the menu list.

– The types are Gross Wt, Net Wt, or Ticket Number.

9. Touch the data box on the screen to the right of

OUTPUT SCALE:,

then select the desired scale required from the menu list.

– The scales choices are All, Active or an individual scale.

– It is recommended to leave the settings for baud, parity, and etc. as shown in

the images in steps 3 and 4.