beautypg.com

Appendix compensation, Error compensation, Open/short compensation – Atec Agilent-16047A User Manual

Page 99: Open/short/load compensation

background image

Error Compensation

Agilent measurement instruments incorporate one of the following four types of error compensation
functions to eliminate residual impedance effects in test fixtures:

1. Open/short Compensation

For a simple measurement system, represented by the equivalent circuit model shown below, resid-
ual impedance values Rs and Ls and admittance values Co and Go can be corrected by:
(i) Measuring open condition for the test fixture’s admittance.
(ii) Measuring short condition for the test fixture’s impedance
(iii)Measuring the test sample, then subtracting the

admittance and impedance.

The above procedure is performed internally by the
open/short compensation.
When the test fixture is directly connected to the mea-
surement instrument, the open/short compensation suf-
ficiently corrects the measurement error.

2. Open/short/load Compensation

When the measurement system is too complicated to be represented as the above equivalent circuit
model, the open/short compensation cannot completely compensate for the residual impedance. In
this case, the open/short/load compensation is used rather than the open/short compensation.
The open/short/load compensation is particularly effective when, for example, the 16065A is used
with the measurement instrument.

The following table lists measurement instruments and available compensation functions.

–: N/A
●: Available
* For more details of the compensation functions, refer to the instruction manuals supplied with the

measurement instrument or Application Note 346-3 (P/N 5091-6553E)

94

Appendix

Compensation

Test Fixture

Instrument

DUT

Model

Open Short Load

Load’s input parameters

compensation

compensation

compensation

4192A

4194A

4263A

R-X (GPIB only)

4263B

Z-Q, R-X, Cp-D, Cp-Q, Cp-G,
Cp-Rp, Cs-D, Cs-Q, Cs-Rs, Ls-D,
Ls-Q, Ls-Rs

4268A

Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D,
Cs-Q, Cs-Rs

4278A

Cp-D, Cp-G

4279A

Cp-D, Cp-G

4284A

same as measurement parameters

4285A

same as measurement parameters

4286A

Rs-Ls

4287A

4288A

Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-Q,
Cs-Rs

4291A/B

Rs-Ls

4294A

Rs-Ls

4338B

4339A/B

E4991A