Up to 3 GHz (7 mm)

SMD

16196U Maintenance Kit

Opt. 16196U-010: Upper electrode, 5 piece set (common
to all 16196x models)
Opt. 16196U-300: Short plate for 0201 (inch)/0603 (mm)
size, 5 piece set (for 16196C)
Opt. 16196U-310: Lower electrode, 5 piece set (for
16196C)

Compensation and Measurement: Open and short
compensations are recommended in combination with the
electrical length compensation before measurement. The
fixture’s electrical length must be entered into the electri-
cal length compensation function of the measurement
instrument first. Next, open compensation is performed
by placing the furnished open plate on top of the insulator
assembly. Short compensation is performed by placing the
furnished shorting plate on top of the insulator assembly.
After performing open and short compensations in combi-
nation with the electrical length compensation, the DUT
is inserted into the test fixture. Once the measurement of
the DUT is complete, remove the DUT from the fixture,
by using the furnished push ring. Refer to the 16196A fig-
ures to see how compensation and measurement is per-
formed.