beautypg.com

Atec Agilent-16047A User Manual

Page 38

background image

42841A Bias Current Source

Dimensions (approx.): 426 (W) x 177 (H) x 498 (D) [mm]

Weight (approx.): 18 kg

DC bias Current Accuracy:

Measurement Configurations:

0-20Adc Bias Configuration (4284A)

Basic Measurement Accuracy: 2% (< 1 kHz), 1% (

1 kHz)

For detailed information, refer to the operation manual of 42841A

0-40Adc Bias Configuration (4284A)

Basic Measurement Accuracy: 2%(< 1 kHz), 1% (

1 kHz)

For detailed information, refer to the operation manual of 42841A

0-10Adc Bias High Frequency Configuration (4285A)

Basic Measurement Accuracy:

f [%] + 4285A's accuracy [%]

f: [MHz]
For detailed information, refer to the operation manual of 42841A
*Opt. 42841A-001 adds SMD test fixture which is available for 42842C.

33

Up to 110 MHz (4-Terminal Pair)

DC Bias Accessories

Description: The 42841A is a bias current source for use
with the 4284A or 4285A. The configured measurement
system has the capability to perform DC bias current
measurements of up to 40 A in the 20 Hz to 1 MHz fre-
quency range and up to 10 A in the 75 kHz to 30 MHz fre-
quency range.
Applicable Instruments: 4284A (Option 4284A-
002)/4285A (Option 4285A-002) only
Maximum DC Bias Current: 20 A
Furnished Accessories:

Options:

42841A-907: Front handle Kit
42841A-908: Rack Mount Kit
42841A-909: Rack Flange and Handle Kit

Description

P/N

Qty.

Bias Current Interface Cable

42841-61640

1

Power Cable

depends on country

1

Operation Manual

42841-90010

1

4284A with Opt. 4284A-002

1 ea.

42841A Bias Current Source

1 ea.

42842A or 42842B Bias Test Fixture

1 ea.

16048A Test Leads

1 ea.

4284A with Opt. 4284A-002

1 ea.

42841A Bias Current Source

2 ea.

42842B Bias Test Fixture

1 ea.

42843A Bias Current Cable

1 ea.

16048A Test Lead

1 ea.

4285A with Opt. 4285A-002

1 ea.

42841A Bias Current Source

1 ea.

42842C Bias Test Fixture*

1 ea.

16048A Test Leads

1 ea.

Range

Resolution Accuracy (±)

0.00 A to 1.00 A

0.01 A

1% + 5 mA

1.1 A to 5.0 A

0.1 A

2%

5.1 A to 20.0 A

0.1 A

3%

20 A measurement system

40 A measurement system

Bias current system configuration for 42842C