Up to 110 mhz (4-terminal pair) – Atec Agilent-16047A User Manual
Page 19
16034E Test Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 2-Terminal
Dimensions (approx.):
128 (W) x 60 (H) x 71 (D) [mm]
Weight (approx.): 270 g
Additional Error:
f: [MHz]
14
Up to 110 MHz (4-Terminal Pair)
SMD
Description: This test fixture is designed for impedance
evaluations of SMD. The minimum SMD size that this fix-
ture is adapted to evaluate is 1.6(L) x 0.8(W) [mm].
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: DC to 40 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended before measurement.
Open compensation is performed by separating the high
and low electrodes from each other. The separation
should be equivalent in size to the DUT’s width. Short
compensation is performed by contacting the high and low
electrodes together. After performing open and short com-
pensations, the DUT is inserted into the test fixture. The
following figures show how compensation and measure-
ment are performed.
Open compensation
Short compensation
Inserting a DUT
Electrode dimensions
Type of Error
Impedance
Proportional Error
±1.5 x (f/10)
2
Description
P/N
Qty.
Operating Manual
16034-90041
1
DUT Dimensions