16034E Test Fixture

Terminal Connector: 4-Terminal Pair, BNC

DUT Connection: 2-Terminal

Dimensions (approx.):

128 (W) x 60 (H) x 71 (D) [mm]

Weight (approx.): 270 g

Additional Error:

f: [MHz]

14

Up to 110 MHz (4-Terminal Pair)

SMD

Description: This test fixture is designed for impedance
evaluations of SMD. The minimum SMD size that this fix-
ture is adapted to evaluate is 1.6(L) x 0.8(W) [mm].
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: DC to 40 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below

Furnished Accessories:

Compensation and Measurement: Open and short
compensations are recommended before measurement.
Open compensation is performed by separating the high
and low electrodes from each other. The separation
should be equivalent in size to the DUT’s width. Short
compensation is performed by contacting the high and low
electrodes together. After performing open and short com-
pensations, the DUT is inserted into the test fixture. The
following figures show how compensation and measure-
ment are performed.

Open compensation

Short compensation

Inserting a DUT

Electrode dimensions

Type of Error

Impedance

Proportional Error

±1.5 x (f/10)

2

Description

P/N

Qty.

Operating Manual

16034-90041

1

DUT Dimensions