23

Up to 110 MHz (4-Terminal Pair)

Other Components

16089A Large Kelvin Clip Leads

Terminal Connector:

4-Terminal Pair, BNC

DUT Connection: 4-Terminal

Cable Length (approx.):

0.94m (from connector to clip's tip)

Weight (approx.): 300 g

Additional Error: The additional error

is negligible when compared to the

instrument's accuracy.

Description: This test fixture makes it possible to mea-
sure odd-shaped components that cannot be measured
with conventional fixtures. It is equipped with two insu-
lated Kelvin clips.
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: 5 Hz to 100 kHz
Maximum Voltage: ±42 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size:
See figure below

Furnished Accessories:

Compensation and Measurement: Open and short
compensations are recommended before measurement.
For open compensation, do not connect the Kelvin clips to
anything. Short compensation is performed by holding a
shorting plate with the Kelvin clips. After performing
open and short compensations, the DUT is held with the
Kelvin clips.

16089B Medium
Kelvin Clip Leads

Terminal Connector: 4-Terminal Pair, BNC

DUT Connection: 4-Terminal

Cable Length (approx.):

0.94m (from connector to clip's tip)

Weight (approx.): 300 g

Additional Error: The additional error is

negligible when compared to the instrument's

accuracy.

Description

P/N

Qty.

Operation and Service Manual

16089-90000

1

Description: This test fixture makes it possible to mea-
sure odd-shaped components that cannot be measured
with conventional fixtures. It is equipped with two insu-
lated Kelvin clips.
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: 5 Hz to 100 kHz
Maximum Voltage: ±42 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below

Furnished Accessories:

Compensation and Measurement: Open and short
compensations are recommended before measurement.
For open compensation, do not connect the Kelvin clips to
aynthing. Short compensation is performed by connecting
the Kelvin clips together. After performing open and short
compensations, the DUT is held with the Kelvin clips.

Description

P/N

Qty.

Operation and Service Manual

16089-90000

1