36

Up to 110 MHz (4-Terminal Pair)

DC Bias Accessories

42842C Bias Current Test Fixture

DUT Connection: 2-Terminal

Bias Voltage Output connector: BNC(f)

Dimensions (approx.):

213(W) x 173(H) x 235(D) [mm]

Weight (approx.): 3100 g

Basic Measurement Accuracy:

f [%] + 4285A's accuracy [%]

f: [MHz]

For detailed information, refer to the operation

manual of 42841A.

P/N 42851-61100

DUT Connection: 2-Terminal

Dimensions (approx.): 122(W) x 60(H) x 58(D)[mm]

Weight (approx.): 145 g

4285A with 42842C and P/N 42851-61100

Description: This test fixture is designed for use with
the 42841A. It is connected directly to the bias current
output terminals of the 42841A. The following features
ensure accurate and safe DC bias measurements:
• Clear plastic cover is provided for safe measurements.
• Opening the cover decreases the voltage generated by

back-e.m.f. (electromotive force) to a level below 40 V
within 0.1 second.

• Heat-sensitive switch prevents DUT from overheating.
• Also equipped with DC bias voltage monitoring
Applicable Instrument: 4285A (Option 4285A-002) only
Frequency: 75 kHz to 30 MHz
Maximum DC Bias Current: 10 A (2 A max for Option
001 SMD Test Fixture)
Operating Temperature: 0°C to 45°C
DUT Size: Smaller than 60(W) x 50(H) x 60(D) [mm]
Furnished Accessories:

Options:

42842C-001: SMD Test Fixture P/N 42851-61100

Compensation and Measurement: Short compensa-
tions is recommended before measurement. Short com-
pensation is performed by shorting the measurement ter-
minals together with the furnished shorting plate. After
compensation, connect the DUT to the measurement ter-
minals and close the test fixture cover.

Description

P/N

Qty.

Shorting Plate

42842-00607

1

Operation and Service Manual

42842-90001

1

Description: This test fixture is supplied with Option
42842C-001 for 42842C. It is designed for the impedance
evaluations of SMD. The minimum SMD size that this
fixture is adapted to evaluate is 1.6(L) x 0.8(W) [mm].
Applicable Instrument: 4285A (Option 4285A-002) only
Frequency: 75 kHz to 30 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Maximum DC Bias Current: 2 A
Operating Temperature: 0°C to 55°C
DUT Size: See figure below

Compensation and Measurement: Refer to the proce-
dure for 16034E.

DUT Dimensions