16043A/B
3-Terminal SMD Test Fixture

Terminal Connector: 4-Terminal Pair, BNC

DUT Connection:

3-Terminal (including the guard terminal)

Dimensions (approx.):

75 (W) x 105 (H) x 95 (D) [mm]

Weight (approx.): 500 g (16043A)

330 g (16043B)

Additional Error:

Guard Terminal Residual L:

6 nH

Guard Terminal Residual R:

20 m

The additional error shown above is the

characteristics of when the DUT connection

is 2-Terminal. Ideally, in a 3-Terminal DUT

connection, the guard effect should be taken

into account in the measurement accuracy.

To acknowledge the guard terminal’s effect,

a characteristics equation is provided in the

operation manual.

20

Up to 110 MHz (4-Terminal Pair)

SMD

Type of Error

Impedance

Proportional Error

0.4 + 7.2 x (f/100)

2

[%]

Open Repeatability

10 + 600 x (f/10) [nS]

Short Repeatability

30 + 6 x (f/10) [m

]

Description: The 16043A/B enable 3-Terminal resonator
SMD to be measured by using the guarding technique.
Having a 4-Terminal Pair configuration it ensures high
precision and repeatability. The 16043A/B accommodates
a wide range of SMD sizes by providing 3 different contact
boards. The 16043A is equipped with a sliding mecha-
nism to enable the measurement of load capacitors in the
3-Terminal resonator. The 16043B is not equipped with
the sliding mechanism. 2-Terminal SMD with bottom
electrodes can be measured as well. As shown in the fig-
ure below, by connecting the device’s G-terminal to instru-
ment’s guard terminal, only Z1 or the resonator will be
measured.

In order to measure Z2 and Z3, the sliding function of
16043A can be utilized.

Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: DC to 110 MHz
Operating Temperature: 0°C to 70˚C
DUT Size: The applicable DUT dimensions are shown
below. A contact board is required for measurement and
its dimensions and ordering info are shown on the next
page.

Guarding technique of an impedance measurement instrument

Electrode configuration of DUT and applicable DUT dimensions