Up to 3 GHz (7 mm)

SMD

Options:

16192A-010: Add EIA/EIAJ industrial standard sized
shorting bar set
16192A-701: Add general sized shorting bar set
16192A-710: Add the magnifying lens and tweezers

Compensation and Measurement: Open and short
compensations are recommended in combination with the
electrical length compensation before measurement. The
fixture’s electrical length must be entered into the electri-
cal length compensation function of the measurement
instrument first. Then open compensation is performed
by separating the high and the low electrodes from each
other. The separation should be equivalent in size to the
DUT’s width. Short compensation is performed by using
option 16192A-010/701 shorting bar set. After performing
open and short compensations in combination with the
electrical length compensation, the DUT is inserted into
the test fixture. The following figures show how compen-
sation and measurement is performed.

Test fixture overview

Electrodes configuration and SMD size