beautypg.com

Up to 3 ghz (7 mm) lead components – Atec Agilent-16047A User Manual

Page 51

background image

46

16092A Spring Clip Fixture

Terminal Connector: 7 mm

DUT Connection: 2-Terminal

Electrical Length: 3.4 mm

Dimensions (approx.): 150(W) x 70(H) x 80(D) [mm]

Weight (approx.): 180 g

Additional Error: See figure below

Up to 3 GHz (7 mm)

Lead Components

Description: This test fixture is designed for impedance
evaluation of both lead and SMD. It is furnished with two
modules that can be readily screwed onto the plate to
measure either lead or SMD.
Applicable Instrument: 4287A, 4294A + 42942A, 4395A
with Opt.4395A-010 + 43961A, 4396B with Opt.4396B-010
+ 43961A, E4991A, (4195A + 41951A, 4286A, 4291A,
4396A)*
When used with 16085B: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: DC to 500 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below

Furnished Accessories:

Compensation and Measurement: Open and short
compensations are recommended in combination with the
electrical length compensation before measurement. The
fixture’s electrical length must be entered into the electri-
cal length compensation function of the measurement
instrument first. When using the SMD module, open
compensation is performed by separating the high and the
low electrodes from each other. The separation should be
equivalent in size to the DUT’s width. Short compensa-
tion is performed by contacting the high and low elec-
trodes together. When using the lead component module,
open compensation is performed by not having the mod-
ule-electrodes be connected to anything. Short compensa-
tion is performed by using the furnished shorting plate.
After performing open and short compensations in combi-
nation with the electrical length compensation, the DUT
is inserted into the test fixture.

Description

P/N

Qty.

Shorting Plate

16092-08010

1

Operating Note

16092-90010

1

Inserting the SMD

Inserting the leaded component

DUT Dimensions