Dc (high resistance) smd – Atec Agilent-16047A User Manual
Page 78

16118A Tweezers Test Fixture
Terminal Connector:
* Interlock Connector enables and disables
the application of source voltage from the
measurement instrument.
DUT Connection: 2-Terminal (with triaxial cable)
Cable Length (approx.): 1 m
Weight (approx.): 410 g
73
DC (High resistance)
SMD
Type
Connector
Input
Triaxial (special screw-type)
Output
High Voltage BNC (special type)
Control
Interlock connector*
Description: The 16118A is designed to operate specifi-
cally with 4339B. The tweezer’s contacts, which makes it
easy to hold SMD. Electrical noise effects are reduced by
the employment of shielded-cables. A built-in interlock-
ing circuit enables safe high-voltage measurements.
Applicable Instruments: 4339B, (4339A)*
* denotes the instrument is obsolete.
Frequency: DC
Maximum Voltage: 100 V
Maximum Current: 0.87 mA
Resistance Measurement Range: 1 x 10
7
to 1 x 10
11
Ω
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open compensation
is recommended before measurement. Separate the elec-
trodes from each other. The separation should be equiva-
lent in size to the DUT’s width. After performing open
compensation, the DUT is sandwiched by the tweezers’
contacts and is measured.
4339B with 16118A
Description
P/N
Qty.
Operation and Service Manual
16118-90010
1
DUT Dimensions