Up to 3 ghz (7 mm) lead components – Atec Agilent-16047A User Manual
Page 53
48
16093B Binding Post Test Fixture
Terminal Connector: 7 mm
DUT Connection: 2-Terminal
Electrical Length: 3.4 mm
Dimensions (approx.): 150(W) x 70(H) x 80(D) [mm]
Weight (approx.): 175 g
Additional Error: See figure below
16194A High Temperature Component Test Fixture
This test fixture is designed for not only lead devices but SMD as well.
See the SMD section for a description of 16194A.
Up to 3 GHz (7 mm)
Lead Components
Description: This test fixture is designed for impedance
evaluation of axial/radial lead type devices. A third bind-
ing post is provided as a guard terminal for three termi-
nal devices.
Applicable Instrument: 4287A, 4294A + 42942A, 4395A
with Opt.4395A-010 + 43961A, 4396B with Opt.4396B-010
+ 43961A, E4991A, (4195A + 41951A, 4286A, 4291A,
4396A)*
When used with 16085B: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: DC to 125 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Operating Temperature: 0°C to 55°C
DUT Size: See figure below
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended in combination with the
electrical length compensation before measurement. The
fixture’s electrical length must be entered into the electri-
cal length compensation function of the measurement
instrument first. Open compensation is performed by not
having the binding posts be connected to anything. Short
compensation is performed by using the furnished short-
ing plate. After performing open and short compensations
in combination with the electrical length compensation,
the DUT is inserted into the test fixture.
Description
P/N
Qty.
Shorting Plate
16092-08010
1
Operating Note
16093-90011
1
Inserting the leaded component
Post spacing