Atec Agilent-16047A User Manual
Page 37
32
Up to 110 MHz (4-Terminal Pair)
DC Bias Accessories
16065A 200Vdc External Voltage
Bias Fixture
Terminal Connector: 4-Terminal Pair, BNC
DUT Connection: 4-Terminal
External Bias Input connector: High Voltage BNC(f)
Dimensions (approx.): 180(W) x 120(H) x 200(D) [mm]
Cable Length (approx.): 40 cm
Weight (approx.): 1500 g
High Voltage BNC(f) connector for external bias input
BNC(f) connector for voltage monitor output
LCR meter with 16065A
16065C 40Vdc External
Voltage Bias Adapter
Terminal Connector: 4-Terminal Pair, BNC
External Bias Input connector: BNC(f)
Dimensions (approx.): 160(W) x 50(H) x 150(D) [mm]
Cable Length (approx.): 210 mm
Weight (approx.): 450 g
LCR meter with 16065C
Description: This test fixture makes it possible to mea-
sure a DUT with up to ±200 V DC bias. The same mod-
ules of 16047A/D can be used to allow measurements of
axial/radial lead components.
Applicable Instruments: 4263B, 4268A, 4279A, 4284A,
4285A, 4288A, 4294A, (4192A, 4194A, 4263A, 4278A)*
* denotes the instrument is obsolete.
Frequency: 50 Hz to 2 MHz
Maximum Voltage: ±200 V peak max (AC+DC)
Blocking Capacitor of 5.6 µF is connected in series with
the Hc terminal.
Operating Temperature: 0°C to 55°C
DUT Size: See the 16047A figure with module sizes.
Furnished Accessories:
Compensation and Measurement: Open, short and
load compensations are recommended before measure-
ment. Short compensation is performed by shorting the
contacts of the test fixture with a shorting plate as
described for 16047A. Load compensation is performed by
inserting a known standard device. After performing
open, short and load compensations, the DUT is connected
to the test fixture.
Description
P/N
Qty.
Module For Axial Lead
16061-70022
1
Module For Radial Lead mounting
16061-70021
1
on fixture
Module For Short Radial Lead
16047-65001
1
Operating Note
16065-90010
1
Description: This adapter is designed to operate specifi-
cally with the 4263B, 4268A and the 4288A. By connect-
ing an external DC voltage source to this adapter, a bias
voltage of up to ±40 V can be supplied to a DUT. The
DUT can be inserted by connecting any direct attachment
4-Terminal Pair test fixture to the adapter.
Applicable Instruments: 4263A*/B, 4268A, 4278A*, 4288A
Frequency: 50 Hz to 1 MHz
Maximum Voltage: ±40 V peak max (AC+DC)
Blocking Capacitor of 50 µF is connected in series with
the Hc terminal.
Operating Temperature: 0°C to 55°C
Applicable Fixtures: 16034E/G/H, 16044A, 16047A/D/E,
16048A/B/D/E, 16089A/B/C/D/E
Furnished Accessories:
Compensation and Measurement: Open and short
compensations are recommended before measurement.
Short compensation is performed by shorting the contacts
of the test fixture that is in use. After performing open
and short compensations, the DUT is connected to the test
fixture.
Description
P/N
Qty.
Operation and Service Manual
16065-90020
1