beautypg.com

H-5223 – Xylem H-5223 User Manual

Page 99

background image

Command:

GOESSTCOLUMNnSHEFCODE=XX

Description:

Set the SHEF code for the n

th

column of self timed GOES data.

Parameters:

‘n’ ranges from 1-20.

Response:

None.

Example:

-> GOESSTCOLUMN3SHEFCODE=DT

5

Command sent to the XL™.

rs

Response from the XL™.

Command:

GOESRRCHANNEL?

Description:

Read the channel to be used for GOES random transmissions.

Parameters:

None.

Response:

An ASCII text string is returned with the channel.

Example:

-> GOESRRCHANNEL?

5

Command sent to the XL™.

GOES RR CHANNEL = 123

rs

Response sent to the host system.

Command:

GOESRRCHANNEL=XXX

Description:

Set the channel for random transmissions of the GOES radio.

Parameters:

The channel is a 3 character decimal value, ranging from 000-200.

Response:

None.

Example:

-> GOESRRCHANNEL=111

5

Command sent to the XL™.

rs

Response sent to the host system.

Command:

GOESRRRATE?

Description:

Read the random transmission rate. This determines the window during which the
GOES radio will transmit after detecting a random trigger.

Parameters:

None.

Response:

An ASCII text string is returned in the format of hh:mm:ss.

C-40 Remote Operation (Command Mode)

H-5223