Parameters:

The parameter string is formatted for two digits used for the hours, (hh); two digits
used for the minutes, (mm); and two digits used for the seconds, (ss). The hour,
minute and second values must be separated with some character as shown above.
Note the space between the command and the parameter.

Response:

None, use the SCANRATE command to see the new value.

Example:

-> SCANRATE=00:01:00

5

Command sent to the XL™.
This sets the scan rate to force a scan once
every minute.

rs

Response sent to the host system.

Command:

NEXTSCAN?

Description:

Get the amount of time until the next scan.

Parameters:

None.

Response:

The response will be an ASCII string in the format of HH:MM:SS indicating the
amount of time until the next scan.

Examples:

-> NEXTSCAN?

5

Command sent to the XL™.

00:14:59

rs

Response sent to the host system.

Command:

NEXTSCANAT?

Description:

Get the scheduled time of the next scan.

Parameters:

None.

Response:

The response will be an ASCII string in the format of HH:MM:SS.

Examples:

-> NEXTSCANAT?

5

Command sent to the XL™.

16:30:00

rs

Response sent to the host system.

Command:

NEXTSCANAT=hh:mm:ss

Description:

Set the time at which the XL™ will performs its next scan.

Parameters:

The format for the time is HH:MM:SS. The time ranges from 00:00:00 to
23:59:59.