Example:
-> SDIVERIFY
5
Command sent to XL™.
11234
rs
Response from XL™.
Command:
SDIMEASURE
Description:
Issue a measure command on the SDI-12 bus using the address specified by
SDITESTADDRESS.
Parameters:
None.
Response:
The XL™ takes the response it receives on the SDI-12 bus and returns it. If no
response is received, the XL™ responds with “Communication Timed Out”. The
response to the measure command is as follows:
atttn
a:
sensor address (“0-9", “A-Z”, “a-z”, “*”, “?”).
ttt:
3 digit integer (000-999) specifying the maximum time, in
seconds, the sensor will take to complete the command and
have measurement data available in its buffer.
n:
single digit integer (0-9) specifying the number of values
that will be placed in the data buffer. If “n” is zero (0), no
data will be available using subsequent “D” commands.
Example:
-> SDIMEASURE
5
Command sent to the XL™.
10022
rs
Response from the XL™.
Command:
SDIDATAn
Description:
Issue a data command on the SDI-12 bus using the address specified by
SDITESTADDRESS.
Parameters:
‘n’ is an integer specifying which data command to issue (DATA0, DATA1, etc).
Response:
The XL™ takes the response it receives on the SDI-12 bus and returns it. If no
response is received, the XL™ responds with “Communication Timed Out”. The
response to the data command is as follows:
+xx.xxx+y.yyy
C-28 Remote Operation (Command Mode)