H-5223 – Xylem H-5223 User Manual
Page 88

x.xxx:
channel A voltage
y.yyy:
channel B voltage
Example:
-> SDIDATA0
5
Command sent to the XL™.
+1.234+9.876
rs
Response from the XL™.
Command:
SDIEXTENDED=<command>
Description:
Issue an extended command on the SDI-12 bus using the address specified by
SDITESTADDRESS.
Parameters:
“command” is the command that is to be issued on the SDI-12 bus. It needs to be a
valid SDI-12 command.
Response:
The XL™ takes the response it receives on the SDI-12 bus and returns it. If no
response is received, the XL™ responds with “Communication Timed Out”. The
response to the extended command depends entirely on the command.
Example:
-> SDIEXTENDED=*A
5
Command sent to the XL™.
1
rs
Response from the XL™.
C.9 Output Commands
Command:
AUTOPRINT?
Description:
Read the auto print flag. If this flag is set, all data being logged also gets sent out
the serial port.
Parameters:
None.
Response:
An ASCII text string is returned with either “On ” or “Off“.
Example:
-> AUTOPRINT?
5
Command sent to the XL™.
Auto Print = Off
rs
Response from the XL™.
Command:
AUTOPRINT=XXX
H-5223
Remote Operation (Command Mode) C-29