beautypg.com

H-5223 – Xylem H-5223 User Manual

Page 77

background image

Command:

SCANNING?

Description:

Reads the scanning flag. This value determines whether or not the XL™ is set to
scan.

Parameters:

None.

Response:

The response will be either the “ON” or “OFF”.

Examples:

-> SCANNING?

5

Command sent to the XL™.

OFF

rs

Response sent to the host system.

Command:

SCANNING=XXX

Description:

Set the scanning flag.

Parameters:

This flag can be set to either “ON” or “OFF”.

Response:

None.

Example:

-> SCANNING=ON

5

Command sent to the XL™

rs

Response from the XL™.

Command

SCANRATE?

Description:

Read the current scan rate interval.

Parameters:

None.

Response:

An ASCII text string is returned with the time formatted as hh:mm:ss. The
response is terminated using a carriage return, line feed.

Example:

-> SCANRATE?

5

Command sent to the XL™.

00:00:30

rs

Response sent to the host system.

Command:

SCANRATE=hh:mm:ss

Description:

Set the time interval for the scanning operation. Keep in mind that the scan rate
should be as fast as, or some multiple faster than the report intervals. See the note
at the end of this section.

C-18 Remote Operation (Command Mode)

H-5223